ICRU supplement to Report 50 that refined the PTV concept: an internal margin for variations in the CTV's size, shape and position, a set-up margin for planning and positioning uncertainties, and a planning organ at risk volume (PRV) around organs at risk. DoseLab has not read the report itself (it is sold by the ICRU); these points are checked as summarized in the open-access Burnet 2004 review. Citation as given in the DEGRO/DGMP statement on ICRU Report 91 (Wilke 2019).
ITV
The internal target volume — a target volume that covers the tumor wherever it moves during treatment, most often over the breathing cycle. NRG lung protocols call the motion-encompassing GTV an IGTV.
What is it?
The internal target volume (ITV) is a target volume that includes the tumor's motion inside the patient — most often the motion caused by breathing.
It comes from ICRU Report 62, which split the margin around the clinical target volume (CTV) into two parts: an internal margin for changes in the CTV's size, shape and position (from respiration or organ filling, for example), and a set-up margin for uncertainties in planning, positioning and beam delivery. The ITV is the CTV with its internal margin; the PTV adds the set-up margin.
Names vary, and the difference matters:
- ITV — in ICRU terms, the motion-encompassing CTV.
- IGTV — the motion-encompassing GTV. AAPM TG-263 defines it as gross disease with a margin for motion; ICTV is the clinical-disease equivalent.
NRG's lung protocol templates use IGTV and explicitly avoid the term ITV "to prevent confusion between a motion encompassing CTV and GTV". Even the image-guidance tables in those same templates say "ITV" loosely, so check which volume a protocol means.
Why does it matter?
If a moving tumor is planned on a single snapshot, part of it can sit outside the high-dose region for some of each breath. An ITV makes the plan cover the whole path instead. The DEGRO working group's definition of lung SBRT lists the ITV concept as the most commonly practiced approach for free-breathing delivery, alongside the mean-target-position concept and real-time tracking, with gating and breath hold as the alternatives that treat only in part of the breathing cycle.
Covering the whole path has a cost. AAPM TG-76 notes that adding margins to cover the full range of tumor motion enlarges the irradiated volume and the amount of healthy tissue receiving high dose — one reason motion-management techniques exist.
The ITV also changes what the PTV margin is for. The NRG lung SBRT template states that its PTV margin is not meant to account for internal motion, because that is already inside the IGTV.
How is it evaluated?
How the volume is built depends on the motion-management technique. The NRG lung SBRT template's guidelines:
- Free breathing with 4DCT (including abdominal compression): the Boolean union of the GTV contoured on every 4DCT phase, or a maximum intensity projection (MIP).
- Gating: the union of the GTV on the gated phases only.
- Breath hold or tracking: a margin for breath-hold or tracking reproducibility is recommended (the template gives a range — use your protocol's).
When a MIP is used, the same template asks that the result be checked against the GTV on each individual phase, and notes that a MIP is not appropriate where the tumor borders soft tissue of similar density, such as the diaphragm or mediastinum. AAPM TG-76 likewise limits the MIP approach to tumors without mediastinal involvement.
Any extra margin for breathing reproducibility, and the PTV margin, come from your protocol and department — DoseLab does not give margin values.
At treatment, the NRG lung templates ask for care to target the ITV accurately with image guidance, and recommend repeat imaging during treatment to verify that the tumor is in the ITV.
Example
The schematic shows a tumor contoured on several breathing phases. The ITV (here, an IGTV) is the envelope that contains every position; the PTV is built around it.
Authoritative resources
Review of the ICRU target-volume concepts. It summarizes ICRU Report 62's internal margin (for variations in the size, shape and position of the CTV, for example from organ filling or respiration) and set-up margin, and its planning organ at risk volume (PRV): a margin around an organ at risk, analogous to the PTV margin, because systematic and random errors apply to organs at risk as much as to the CTV. It notes that a PRV substantially enlarges the organ and may create dilemmas with target dose, so it is needed only in certain circumstances — around organs whose damage is especially dangerous, such as the spinal cord, a "serial" tissue. Full text read in Europe PMC (PMC1434601).
AAPM task group report on respiratory motion in thoracic, abdominal and pelvic treatment. It groups management approaches into motion-encompassing methods, respiratory gating, breath-hold techniques, forced shallow breathing (such as abdominal compression) and respiration-synchronized (tracking) techniques. Read in the AAPM report PDF (Report No. 91, July 2006): using ICRU Report 62 nomenclature, the PTV adds margins to the CTV for respiratory (intrafraction) motion, interfraction motion and set-up error; without a respiratory management device the entire range of motion should be considered in the internal margin; and for lung tumors a maximum intensity projection (MIP) can give the motion-encompassing volume, provided there is no mediastinal involvement.
Recommendations for standardized naming of targets, organs at risk and dose-volume metrics. Section 9 sets out the DVH-metric nomenclature DoseLab follows: Vx (volume receiving at least dose x), Dx (minimum dose to the hottest sub-volume x), CVx and DCx (the "cold" complements), output units in square brackets (Gy or % of prescription; cc or % of the structure), and an EQD2Gy unit with the α/β written in parentheses — for example D90%(10)[EQD2Gy]. Section 10 asks that DVH metrics say whether a lower-dose PTV is segmented (excludes the higher-dose PTV) or not. Read in the AAPM report PDF.
The template NRG Oncology gives investigators for writing the radiation section of a lung SBRT trial. Its imaging section allows PET/CT or MRI for structure definition, asks that the images be registered accurately to the planning scan (following AAPM TG-132 for deformable registration), notes that PET may be needed to separate tumor from atelectasis or fibrosis, and suggests an extra margin when a fused image is the primary image for the GTV. Its target and planning sections (5.2.4 and 5.2.6–5.2.8) define the plan-quality measures for lung SBRT trials: the prescription as the dose covering 95% of the PTV, a D99% requirement, R100% (prescription isodose volume ÷ PTV), R50% (50% prescription isodose volume ÷ PTV) and D2cm (maximum dose at 2 cm from the PTV in any direction, as a percentage of the prescription), with a table of allowable dose spillage by PTV volume. It describes a 20 mm PTV expansion structure used to control intermediate dose spillage, asks that dose above 105% of prescription be avoided outside the PTV, describes a deliberately heterogeneous dose with the hotspot in the central GTV, defines the maximum dose on a 0.03 cc volume, and ranks planning priorities when target coverage and organ limits conflict.
The NRG Oncology template for the radiation section of a conventionally fractionated lung trial. Its imaging section allows PET/CT or MRI for structure definition but asks that the final contours be confirmed on the planning CT, follows AAPM TG-132 for deformable registration, and notes that PET can help separate tumor from atelectasis. For plan evaluation it defines the prescription as the dose covering 95% of the PTV, with a protocol-specific D99% requirement, and asks that any dose above 105% of the prescription fall within the PTV rather than in normal tissue outside it.
DEGRO working-group statement defining SBRT for stage I NSCLC. It lists the motion-management approaches in routine use — free-breathing delivery with the internal target volume concept (described as the most commonly practiced), the mean target position concept or real-time tracking, and delivery in specific breathing phases by gating or breath hold — and names safety margins for critical organs (the PRV concept), repositioning or replanning as ways to avoid organ overdose. Full text read in Europe PMC (PMC3889283).